FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS® 4600 Chemistry System, Product Code: 6802445 Product Usage: For use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immuno rate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

Recall number
Z-1469-2012
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS® 4600 Chemistry System, Product Code: 6802445 Product Usage: For use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immuno rate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Status
Terminated
Initiated
2012-03-15
Posted
2012-04-26
Terminated
2017-09-20
Reason
Ortho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
Root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics (OCD) sent an Important Product Correction Notification letter dated March 15, 2012 via FedEx overnight courier to all affected customers. Foreign affiliates were informed by email on 15 March 2012 of the issue and instructed to notify their consignees of the issue and actions. The letter identified the affected products, problem and actions to be taken. The letter instructed customers to: 1) Consider performing the backup procedure more frequently to help prevent the loss of data on your system; 2) Contact OCD Customer Technical Service representatives if a hard drive failure occurs on your system; A Field Engineer will be dispatched to replace the hard drive on your system; 3) Complete and return the Confirmation of Receipt form upon receipt of this notification; and 4) Post this notification by each VITROS System in your facility or with the user documentation. For questions, refer to the Questions and Answers section. For additional questions, contact Customer Technical Services at 1-800-421-3311.
Quantity
50
Distribution
Worldwide Distribution
Lot, serial or model codes
Serial number range: 46000108 - 46000157
510(k) numbers
["K031924"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000305841
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register