Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 15, 140 mm stem length, tivanium TI-6AL-4V alloy, sterile; Cat. no. 7861-15-04 (00786101504).

Recall number
Z-1469-06
Recalling firm
Zimmer Inc.
Product
Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 15, 140 mm stem length, tivanium TI-6AL-4V alloy, sterile; Cat. no. 7861-15-04 (00786101504).
Status
Terminated
Initiated
2006-08-02
Posted
2006-08-31
Terminated
2007-05-04
Reason
Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly.
Root cause
Other
Action
Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand.
Quantity
181 of all devices
Distribution
Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland.
Lot, serial or model codes
Lots 60454201, 60454202, 60454203, 60454204, 60465019 and 60469053.
510(k) numbers
["K961378"]
PMA numbers
not on file
Product code
LWJ
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register