Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.

Recall number
Z-1465-2013
Recalling firm
Biomet, Inc.
Product
Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.
Status
Terminated
Initiated
2013-04-26
Posted
2013-06-05
Terminated
2013-08-29
Reason
Biomet is recalling Active Articulation E1 Hip Bearing implants. Biomet has initiated this action following an investigation which identified that outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. The implant itself is etched with the correct O.D. size.
Root cause
Labeling mix-ups
Action
Biomet sent an Urgent Medical Device Recall letter dated April 26, 2013 to consignees detailing that the outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. This action requires the immediate location and discontinued use of the product and its return to Biomet. Ciustomers were advised to: Immediately locate and remove the identified device(s) listed from circulation . -Carefully follow the instructions on the enclosed "FAX Back Response Form" . - Fax a copy of the Response Form to 574-372-1683 prior to return of product. -Use priority carrier for your shipment. - If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear O.R. Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products.
Quantity
8 units
Distribution
Distributed in New York.
Lot, serial or model codes
Part numbers: EP-200162, EP-200162, EP-200164, and EP-200166 with Lot numbers: 027050, 101470, 115950, 101480, 247990, 948810, 967270, 027070, 287820, 729870, and 860880.
510(k) numbers
["K101336"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register