FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Recall number
- Z-1464-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Status
- Open, Classified
- Initiated
- 2026-01-08
- Posted
- 2026-02-24
- Terminated
- not on file
- Reason
- Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
- Root cause
- Process control
- Action
- On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.
- Quantity
- 2020
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FKG
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28