Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.

Recall number
Z-1464-2014
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.
Status
Terminated
Initiated
2014-03-06
Posted
2014-04-16
Terminated
2014-11-05
Reason
A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
Root cause
Software in the Use Environment
Action
.The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated 2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Quantity
116 units
Distribution
Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.
Lot, serial or model codes
System code: 728323; Serial #: 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320031, 320032, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047, 320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320074, 320075, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320085, 320086, 320087, 320088, 320089, 320090, 320092, 320093, 320095, 320096, 320097, 320098, 320099, 320101, 320102, 320103, 320104, 320105,320106, 320107, 320109, 320110, 320111, 320112, 320114, 320115, 320116, 320117, 320118, 320119, 320120, 320121, 320122, 336013 & 336015.
510(k) numbers
["K033326"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register