Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics: RapidLab 1200 Systems, Model 1245 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC, Diagnostics Division

Recall number
Z-1464-06
Recalling firm
Siemens Healthcare Diagnostics
Product
RapidLab 1200 Systems, Model 1245 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC, Diagnostics Division
Status
Terminated
Initiated
2006-04-04
Posted
2006-08-29
Terminated
2012-07-30
Reason
Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data management system.
Root cause
Software design
Action
On 4/04/06, Bayer HealthCare LLC (Diagnostics Division) sent Support Bulletins to all affected Bayer HealthCare LLC Branches via e-mail in order to inform them of the issue, and to provide them with instructions for the interim work-around until the new software version is installed. Customer Bulletins were prepared and sent to the Branches both in the United States and globally for communication with the affected customers.
Quantity
Domestic - 124 units, Foreign-364 units (all systems)
Distribution
Worldwide
Lot, serial or model codes
Part No. 05061537; all units using Version 1.0, 1.01, or 1.02 software connected to LIS, Rapidlink¿, or Rapidcomm™ data management systems.
510(k) numbers
["K031560"]
PMA numbers
not on file
Product code
CHL
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register