Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm

Recall number
Z-1463-2018
Recalling firm
Integra LifeSciences Corp.
Product
The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm
Status
Terminated
Initiated
2017-12-20
Posted
2018-04-20
Terminated
2020-02-24
Reason
Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required.
Root cause
Device Design
Action
Firm sent letters to consignees on December 20, 2017. Firm asked consignees to examine inventory and determine if consignee had affected product. Firm promised to send new stem trials along with a return shipping label. Firm asked to complete the Acknowledgment and Return Form and email or fax back to firm. For further questions, please call (609) 275-0500.
Quantity
1619 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution.
Lot, serial or model codes
TRL092002506  Lot Numbers: 19419-1 15916-1 15457-1 201274-1 201893-1 2046177  TRL092002507 Lot Numbers: 19419-2 15916-2 15457-2 201274-2 201893-2 2046178  TRL092002508 Lot Numbers: 19419-3 15916-3 15457-3 201274-3 201893-3 2046179  TRL092002509 Lot Numbers: 19419-4 15916-4 15457-4 201274-4 201893-4 20461710  TRL092002510 Lot Numbers: 19419-5 15916-5 15457-5 201274-5 201893-5 20461711  TRL092002511 Lot Numbers: 19419-6 15916-6 15457-6 201274-6 201893-6 20461712  TRL092002512  Lot Numbers: 19419-7 15916-7 15457-7 201274-7 201893-7 20461713  TRL092002513 Lot Numbers: 19419-8 15916-8 15457-8 201274-8 201893-8 20461714  TRL092002514  Lot Numbers: 19419-9 15916-9 15457-9 201274-9 201893-9 20461715  TRL092002515 Lot Numbers: 19419-10 15916-10 15457-10 201274-10 201893-10 20461716  TRL092002516 Lot Numbers: 19419-11 15916-11 15457-11 201274-11 201893-11 20461717
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWT
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28