Laax, Inc.: TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).

Recall number
Z-1462-2015
Recalling firm
Laax, Inc.
Product
TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).
Status
Terminated
Initiated
2015-03-25
Posted
2015-04-23
Terminated
2017-03-17
Reason
Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure of the TigerPaw System II fastener.
Root cause
Unknown/Undetermined by firm
Action
The firm, LAAx, Inc.on behalf of Maquet Medical Systems, sent an "URGENT MEDICAL DEVICE RECALL (Removal) IMMEDIATE ACTION REQUIRED " letter dated March 30, 2015 to all affected sites by express mail. The letter describers the product, problem and actions to be taken. The customers were instructed to examine your inventory immediately to determine if you have any TIGERPAW System II products; remove the products, quarantine them and place in a secure location; complete and return the MEDICAL DEVICE RECALL (Removal) RESPONSE FORM by e-mailing a scanned copy to:Tigerpaw2015@maquet.com or FAX to: 1-(973)-396-3607. If you have acted as a distributor for the TIGERPAW System product II, immediately contact those accounts, advise them of the recall situationa and have them return all stock to you. Your Marquet Medical System USA Sales Representative will contact you to assist with the return of any product. If you have any additional questions, contact your local Maquet Medical Systems USA Sales Representative or our Customer Service at 1-888-880-2874 (6:00 AM - 5:00 PM PST).
Quantity
~4,154 units distributed in US - both part numbers
Distribution
Worldwide Distribution: US (nationwide) and countries of:Australia, Bahrain, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy Netherland, Norway, Poland, Qatar, Slovenia, Spain, Saudi Arabia, Sweden, Switzerland, Turkey and United Kingdom.
Lot, serial or model codes
Part number C-TP-1509 (9 connector)  All serial numbers/lots
510(k) numbers
["K111064"]
PMA numbers
not on file
Product code
GDW
Firm FEI
3008788191
Firm city
Livermore
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register