Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: GEM NEUROTUBE, Absorbable Woven Polyglycolic Acid Mesh Tube, Product Codes: a) 5311-01240-010, b) 5311-01240-012; for peripheral nerve repair

Recall number
Z-1461-2023
Recalling firm
Baxter Healthcare Corporation
Product
GEM NEUROTUBE, Absorbable Woven Polyglycolic Acid Mesh Tube, Product Codes: a) 5311-01240-010, b) 5311-01240-012; for peripheral nerve repair
Status
Open, Classified
Initiated
2023-03-27
Posted
2023-04-25
Terminated
not on file
Reason
The product is brittle and potentially crumbling upon handling or when being removed from its package.
Root cause
Under Investigation by firm
Action
Baxter issued an Urgent Medical Device Recall notice to its consignees on 03/27/2023 by mail (USPS). The notice explained the issue and the risk and requested the following actions: 1. Immediately locate, isolate, and cease all use of the affected lot number of the product. 2. Contact Synovis MCA to arrange for return and replacement of the product. Synovis MCA can be reached at 800-510-3318 or 205-941-0111 between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. Please have your Synovis MCA account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any inventory. Log in to the portal using the account number listed on the enclosed reply form instruction sheet. 4. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them.
Quantity
169 units
Distribution
Worldwide distribution - US Nationwide and the countries of China, Germany, Sweden, Norway, Netherlands, Greece, Switzerland, France
Lot, serial or model codes
a) 5311-01240-010, Lot Number 19012112, exp. 12/31/2023; b) 5311-01240-012, Lot Number 19012112, exp. 12/31/2023
510(k) numbers
["K983007"]
PMA numbers
not on file
Product code
JXI
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register