Intuitive Surgical, Inc.: Da Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier, Model 420327 Da Vinci S/Si Small Clip Applier, Model 420003

Recall number
Z-1459-2022
Recalling firm
Intuitive Surgical, Inc.
Product
Da Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier, Model 420327 Da Vinci S/Si Small Clip Applier, Model 420003
Status
Open, Classified
Initiated
2022-07-01
Posted
not on file
Terminated
not on file
Reason
Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.
Root cause
Nonconforming Material/Component
Action
On 07/01/2022, Intuitive Surgical sent an "Urgent Medical Device Correction" Letter to customers via FedEx courier informing them there is the potential for unexpected motion while using the da Vinci S/Si and X/Xi Endowrist Clip Applier Instruments. This can happen when one of the da Vinci S/Si and X/Xi Clip Applier grip discs are disengaged from the sterile adapter disc during the engagement routine that occurs prior to insertion of the instrument into the cannula and into the patient. To mitigate this issue, Intuitive Surgical has developed a Clip Applier Manual Self-Test (Appendix A) that is part of the Urgent Medical Device Correction Letter. This test is performed by moving the Clip Applier grips through the full range of yaw motion (side-to-side in the plane of the instrument grips) and observing the movement of both the instrument grips prior to placing a clip around a vessel/tissue. This test should be performed in an open space to prevent any collisions. Pass Criteria " During the yaw movement, if the grips remain open with no observation of grips closing then the instrument passes the self-test and users may proceed with using the clip applier instrument in a procedure. Fail Criteria " During yaw movement, if the grips begin to close as opposed to the grips remaining open then there is potential that one of the grip discs is not engaged. Instrument should be removed, reinstalled, and inspected for engagement. If the self- test fails, remove Clip Applier instrument, re-install and repeat self-test. If it fails again, use a different Clip Applier instrument and return the instrument using the standard Intuitive RMA process. Please call customer service to return the affected product using the appropriate local number listed in section 6 of the letter. Credit will be issued when the reported unexpected motion is confirmed. For questions: " US, Puerto Rico and Dominican Republic: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg
Quantity
62,895 Clip appliers
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Lot, serial or model codes
Model/Model Number/UDI Number Da Vinci S/Si Large Clip Applier / 420230 / 00886874111680 Da Vinci S/Si Medium-Large Clip Applier / 420327 / 00886874111826 Da Vinci S/Si Small Clip Applier / 420003 / 00886874111161 Serial Numbers: All
510(k) numbers
["K050369"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register