Skip to the content

FDA Medical Device Recalls (openFDA)

Braemar Manufacturing, LLC: Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

Recall number
Z-1459-2020
Recalling firm
Braemar Manufacturing, LLC
Product
Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only
Status
Terminated
Initiated
2020-01-22
Posted
not on file
Terminated
2022-05-27
Reason
An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.
Root cause
Software design
Action
Consignees notified via email with recall letter dated 01/22/2020. The letter identified affected product, stated reason for recall, and provided instructions on how to reset the error by following provided instructions, or to send the equipment to Braemar for repair.
Quantity
10582
Distribution
Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.
Lot, serial or model codes
All devices manufactured or distributed prior to 01/01/2020 are affected.
510(k) numbers
["K071733"]
PMA numbers
not on file
Product code
MWJ
Firm FEI
1000520785
Firm city
Eagan
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register