Braemar Manufacturing, LLC: Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only
- Recall number
- Z-1459-2020
- Recalling firm
- Braemar Manufacturing, LLC
- Product
- Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only
- Status
- Terminated
- Initiated
- 2020-01-22
- Posted
- not on file
- Terminated
- 2022-05-27
- Reason
- An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.
- Root cause
- Software design
- Action
- Consignees notified via email with recall letter dated 01/22/2020. The letter identified affected product, stated reason for recall, and provided instructions on how to reset the error by following provided instructions, or to send the equipment to Braemar for repair.
- Quantity
- 10582
- Distribution
- Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.
- Lot, serial or model codes
- All devices manufactured or distributed prior to 01/01/2020 are affected.
- 510(k) numbers
- ["K071733"]
- PMA numbers
- not on file
- Product code
- MWJ
- Firm FEI
- 1000520785
- Firm city
- Eagan
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28