Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Molecular Systems, Inc.: COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD

Recall number
Z-1459-2011
Recalling firm
Roche Molecular Systems, Inc.
Product
COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD
Status
Terminated
Initiated
2010-12-28
Posted
2011-03-01
Terminated
2012-08-27
Reason
There is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument.
Root cause
Employee error
Action
Roche sent Urgent Medical Device Correction Letters to their consignees on December 28, 2010 via Fed Ex. The letters advise the customers to verify the correct configuration of the CSI reagent cassette, remind customers to follow the package insert instructions for including all required controls for each batch and advise that review of previously generated results may be warranted. A UMDC Faxback form was attached for the consignee to complete. For questions regarding this recall call 908-253-7569.
Quantity
93 kits to US; 22,119 kits to foreign
Distribution
Worldwide Distribution - USA including CA, CT, IN, MI, MN, NC, NE, NY, OR, TX, WA, and WI and the countries of Australia, Austria, Azerbaijan, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Egypt, Finland, France, Germany, Ghana, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mauritania, Mongolia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Singapore, Slovakia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and the United Kingdom
Lot, serial or model codes
Catalog number 05027012190, 05393850190, 05393850190, 05393868190 Lot #K11236, K15440, L00846, M05135, M11190, M12625, M14232, N00159, N02341, N05553, N05554, N07543, N08567
510(k) numbers
not on file
PMA numbers
not on file
Product code
MKT
Firm FEI
2243471
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register