Philips Medical Systems North America Co. Phillips: The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030 19641-SW R 2.5.3 9896 030 19861-SW R 2.6.1 9896 030 19862-SW R 2.6.1 9896 030 19863-SW R 2.6.3

Recall number
Z-1458-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030 19641-SW R 2.5.3 9896 030 19861-SW R 2.6.1 9896 030 19862-SW R 2.6.1 9896 030 19863-SW R 2.6.3
Status
Terminated
Initiated
2010-02-05
Posted
2011-03-01
Terminated
2011-08-10
Reason
In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of additional X-RAY during treatment.
Root cause
Other
Action
Philips began sendng out the URGENT-Device Correction Notice letter dated February 5, 2010, ro all affected consignees on February 10, 2010. The consignees were advised to change the patient position from "Prone" immediately to "Supine" and store the corrected ExamCard in the hospital database if they use the ExamCard, MobiFlex dual Injection. Consignees were instructed to call Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and also reference "FCO 78100318" with any communication.
Quantity
82 units
Distribution
Nationwide Distribution including the following states: AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV.AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV. There was no foreign consignees.
Lot, serial or model codes
Site Numbers: 533860, 533900, 535230, 535664, 538574, 539374, 540482, 541065, 541851, 542236, 544039, 544148, 544696, 545034, 545732, 545981, 546206, 547843, 547897, 548040, 548064, 549389, 549909, 550324, 550439, 551025, 551583, 552453, 552526, 552609, 553116, 554715, 556146, 556161, 556895, 557599, 41443708, 41444456, 41444458, 41444745, 41445015, 41445018, 41445530, 41445594, 41445595, 41446119, 41455905, 41455908, 41455913, 41455915, 41455923, 41455938, 41455940, 41455954, 42047935, 42057444, 42293607, 42392891, 42608348, 42850814, 42892497, 42917294, 43708886, 44062048, 44247559, 44605983, 45130117, 45245963, 45246494, 45246504, 45445404, 47157157, 47230785, 47455618, 47480028, 47848863, 49060365, 49060404, 49416204, 49424702, 49776644,  and 49861074.
510(k) numbers
["K041602"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register