Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes, Inc.: Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.

Recall number
Z-1456-2014
Recalling firm
Synthes, Inc.
Product
Synthes CMF Battery Powered Driver, Part no. 305.840. Used to place screws in the cranialfacial bones, and to drill and/or tap holes to prepare cranialfacial bones for screw insertion.
Status
Terminated
Initiated
2014-01-30
Posted
2014-04-14
Terminated
2016-08-26
Reason
The CMF Battery Powered Driver graphic case contains outlines of the Battery Powered Driver, taps, screwdriver blades, and a screw holding sleeve. However, the device is not intended to be used in conjunction with a tap for tapping holes in a patient's skull, upper and/or lower jaw. Therefore some outlines in the graphic case suggest a potential use of this device that is considered off-label.
Root cause
Labeling design
Action
Synthes sent an Urgent Notice: Field Safety Notification letter dated January 30, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please take the following actions; Please see the Battery Powered Driver Users Manual for on-label procedures. You can access an electronic version of the Battery Powered Driver Users Manual by accessing the following : linlc http://www.synthes.com/pages/default.aspx. Click on North America Tab at the top of the screen, and then type Battery Powered Driver Users Manual in the search bar to download the Users Manual Forward this Field Safety Notification to anyone in your facility that needs to be informed. If the graphic case has been forwarded to another facility, contact that facility. " Maintain a copy of this notice. If you have any questions, please call 610-719-5450 or contact your Synthes Sales Consultant. Thank you for your attention to this issue.
Quantity
20081
Distribution
Nationwide Distribution
Lot, serial or model codes
Part no. 305.840, with lot nos.: 5680577, 5701123, 5713215, 5763130, 5783316, 5787131, 5828296, 6124867, 6136125, 6150803, 6167974,  6167975, 6218358, 6222322, 6258395, 6365430, 6365431, 6390225,  6425013, 6456634, 6487035, 6556704, 6624880, 6650136, 6668870,  6743859, 6844968, 6926708, 6983799, 7008819, 7038706, 7221395,  7138319.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GXL
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register