Olympus Corporation of the Americas: Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications.

Recall number
Z-1455-2026
Recalling firm
Olympus Corporation of the Americas
Product
Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications.
Status
Open, Classified
Initiated
2025-11-11
Posted
2026-02-20
Terminated
not on file
Reason
Complaints of the ceramic tip of the resection sheath breaking have been received.
Root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at Cynthia.Ow@Olympus.com or by phone at 947-999-3203.
Quantity
0 units
Distribution
US Nationwide Distribution.
Lot, serial or model codes
Model No. A2641; UDI: 04042761004060; All Lots.
510(k) numbers
["K931995"]
PMA numbers
not on file
Product code
FJL
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register