FDA Medical Device Recalls (openFDA)
Abbott Laboratories: REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
- Recall number
- Z-1455-2022
- Recalling firm
- Abbott Laboratories
- Product
- REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
- Status
- Open, Classified
- Initiated
- 2022-06-09
- Posted
- not on file
- Terminated
- not on file
- Reason
- There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
- Root cause
- Under Investigation by firm
- Action
- On June 9, 2022, Abbott issued a "Urgent Medical Device Correction" to affected consignees via FedEx. 1. Immediately install the updated assay file versions listed below (as applicable for your laboratory), and manually configure the Low-Linearity value. The assay files can be obtained from www.corelaboratory.abbott 2. For the whole blood application, manually configure the THbWB Low-Linearity to 53.8278. For the hemolysate application, manually configure the THbH Low-Linearity to 1250. For detailed information on editing assay parameters, refer to Changing assay configuration settings, Change a linearity range in the ARCHITECT System Operations Manual, Section 2. 3. Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for reviewing previously reported patient results. 4. Complete and return the Customer Reply Form. 5. Please retain this letter for your laboratory records. 6. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter.
- Quantity
- 7,391 units
- Distribution
- US: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NJ NV NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: ALBANIA ALGERIA ANDORRA ANGOLA ARGENTINA ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BANGLADESH BELARUS BELGIUM BERMUDA BOSNIA AND HERZEGOVI BRAZIL BRUNEI BULGARIA BURKINA FASO CANADA CHILE CHINA COLOMBIA COSTA RICA Country CROATIA CYPRUS CZECH REPUBLIC ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GABON GERMANY GREECE GUATEMALA HONG KONG HUNGARY INDIA INDONESIA IRAQ IRELAND ISRAEL ITALY JAMAICA JORDAN KAZAKHSTAN KENYA KUWAIT KYRGYZSTAN LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MALDIVES MALI MEXICO MOLDOVA MOROCCO MOZAMBIQUE NAMIBIA NETHERLANDS NIGERIA NORWAY OMAN PAKISTAN PALESTINIAN TERRITOR PANAMA PARAGUAY PERU PHILIPPINES POLAND PORTUGAL QATAR REPUBLIC OF KOSOVO REPUBLIC OF YEMEN ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SPAIN ST. LUCIA ST. VINCENT SWEDEN SWITZERLAND TANZANIA THAILAND THE GAMBIA TRINIDAD & TOBAGO TUNISIA TURKEY TURKS & CAICOS UAE UGANDA UNITED KINGDOM URUGUAY UZBEKISTAN VIETNAM ZIMBABWE
- Lot, serial or model codes
- UDI/DI: 00380740135591/ Lots: 59561UQ04, 59779UQ05, 60229UQ07, 60540UQ09, 61011UQ10, 61349UQ01, 61658UQ01
- 510(k) numbers
- ["K130255"]
- PMA numbers
- not on file
- Product code
- PDJ
- Firm FEI
- 3016438761
- Firm city
- Irving
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28