Skip to the content

FDA Medical Device Recalls (openFDA)

Micro Therapeutics Inc, Dba Ev3 Neurovascular: Alligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.

Recall number
Z-1454-2014
Recalling firm
Micro Therapeutics Inc, Dba Ev3 Neurovascular
Product
Alligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.
Status
Terminated
Initiated
2014-04-01
Posted
2014-04-16
Terminated
2014-10-31
Reason
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
Root cause
Process change control
Action
A customer notification letter was sent on 4/1/14 to all customers who purchased the Pipeline Embolization Device and the Alligator Retrieval Device. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the customer response form and fax the completed form to Covidien at (763) 591-3519. Customers with questions are instructed to contact Dr. Mark A. Turco at (202) 310-5120. A press release was sent on 4/11/14 to Business Wire Newsroom to inform the public about the recall from Covidien. The recall is regarding certain lots of its Pipeline Embolization Device and Alligator Retrieval Device where the polytetrafluoroethylene (PTFE) coating applied to the delivery wire could delaminate adn detach from the devices. Consumers are instructed to report any adverse reactions or quality problems experienced with the use of the product to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Consumers with may also contact Covidien Customer Service at (800) 716-6700 between 7am to 7pm (central) or email at customerserviceus@covidien.com.
Quantity
567 units total (323 units in US)
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Germany, France, UK, Turkey, Argentina, Canada, Colombia, Mexico, , Austria, Belgium, Denmark, Hungary, Italy, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, Slovenia, Spain, Switzerland, and UAE.
Lot, serial or model codes
ARD Model/ Lot  FA8881040 / 9769118  FA8881030 / 9816048 FA8881040 / 9754682  FA8881050 / 9772283  FA8881040 / 9752800 FA8881030 / 9751662  FA8881030 / 9773525  FA8881020 / 9832735 FA8881040 / 9757667  FA8881020 / 9774620  FA8881030 / 9834039 FA8881040 / 9758530  FA8881020 / 9774620  FA8881030 / 9834039 FA8881030 / 9758532  FA8881030 / 9783730  FA8881020 / 9836553 FA8881050 / 9760946  FA8881040 / 9786013  FA8881030 / 9834039 FA8881030 / 9761690  FA8881040 / 9787928  FA8881050 / 9848544 FA8881020 / 9761692  FA8881030 / 9791385  FA8881030 / 9848543 FA8881030 / 9758532  FA8881040 / 9793305  FA8881030 / 9853152 FA8881040 / 9764206  FA8881030 / 9798648  FA8881030 / 9854769 FA8881040 / 9764206  FA8881030 / 9798648  FA8881030 / 9858153 FA8881040 / 9764212  FA8881050 / 9801041  FA8881030 / 9865759 FA8881030 / 9766949  FA8881020 / 9811215  FA8881050 / 9868025 FA8881030 / 9769117  FA8881030 / 9813908
510(k) numbers
["K043580"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3004904811
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register