Skip to the content

FDA Medical Device Recalls (openFDA)

Symbios Medical Products, LLC: Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated catheter & BIOPATCH; Go Pump Kit 510349-BP, Epidural Catheter, BioPatch; Go Pump Kit 510642 Disposable Single Infusion Pump, Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. The medication is delivered via tubing to the surgical site at a very slow rate

Recall number
Z-1454-2013
Recalling firm
Symbios Medical Products, LLC
Product
Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated catheter & BIOPATCH; Go Pump Kit 510349-BP, Epidural Catheter, BioPatch; Go Pump Kit 510642 Disposable Single Infusion Pump, Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. The medication is delivered via tubing to the surgical site at a very slow rate
Status
Terminated
Initiated
2013-05-10
Posted
2013-06-06
Terminated
2014-03-07
Reason
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Root cause
Under Investigation by firm
Action
Symbios sent UrgentMedical Device Recall notices to all consignees beginning May 10, 2013. A second corrected/amended notice was sent on May 14, 2013 to consignees and a National Press release was issued. May 30,2013, a third amended Urgent Recall Notice was sent to all consignees correcting the pump description and lot number range table on page 5 of the notice. The notices and press release provided identification of the recalled devices including devices in the previous February 2013 recall, described possible adverse events, and included directions to to quarantine product in a secure location, in preparation for return. Consignees were asked to call 317 225-4447 ext. 25, 8:00 a.m to 4:00 p.m. EDT to receive a Returned Goods Authorization Number prior to returning product. Consignees were asked to complete and return the Attached Verification Form. Question and Concerns should be addressed to 317-225-4447 ext 25.
Quantity
all kits 33, 543
Distribution
US Nationwide distribution.
Lot, serial or model codes
Pump Part 510180, Lot 11-100453 to 12-101003 found in:  Go Pump Kit 510201-BP. Lot 12-100108 to 12-101538  Go Pump Kit 510349-BP. Lot 12-100935 Go Pump Kit 510642, Lot 12-100616 to 13-100590
510(k) numbers
not on file
PMA numbers
not on file
Product code
MEB
Firm FEI
3005203102
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source