Skip to the content

FDA Medical Device Recalls (openFDA)

Micro Therapeutics Inc, Dba Ev3 Neurovascular: Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.

Recall number
Z-1453-2014
Recalling firm
Micro Therapeutics Inc, Dba Ev3 Neurovascular
Product
Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
Status
Terminated
Initiated
2014-04-01
Posted
2014-04-16
Terminated
2014-10-31
Reason
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
Root cause
Process change control
Action
A customer notification letter was sent on 4/1/14 to all customers who purchased the Pipeline Embolization Device and the Alligator Retrieval Device. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the customer response form and fax the completed form to Covidien at (763) 591-3519. Customers with questions are instructed to contact Dr. Mark A. Turco at (202) 310-5120. A press release was sent on 4/11/14 to Business Wire Newsroom to inform the public about the recall from Covidien. The recall is regarding certain lots of its Pipeline Embolization Device and Alligator Retrieval Device where the polytetrafluoroethylene (PTFE) coating applied to the delivery wire could delaminate adn detach from the devices. Consumers are instructed to report any adverse reactions or quality problems experienced with the use of the product to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Consumers with may also contact Covidien Customer Service at (800) 716-6700 between 7am to 7pm (central) or email at customerserviceus@covidien.com.
Quantity
32 units total (23 units in US)
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Germany, France, UK, Turkey, Argentina, Canada, Colombia, Mexico, , Austria, Belgium, Denmark, Hungary, Italy, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, Slovenia, Spain, Switzerland, and UAE.
Lot, serial or model codes
PED Model/ Lot  FA7735014 / 9879718  FA7742512 / 9864827 FA7150035 / 9855921 FA7735016 / 9856429  FA7742518 / 9856431 FA7730010 / 9869991  FA7737514 / 9855933  FA7742518 / 9860762 FA7732514 / 9873854  FA7737516 / 9858010  FA7745014 / 9874111 FA7735014 / 9855265  FA7740016 / 9859214  FA7745016 / 9879736 FA7735014 / 9875267  FA7740016 / 9879180  FA7747514 / 9868888
510(k) numbers
not on file
PMA numbers
["P100018"]
Product code
DQY
Firm FEI
3004904811
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register