FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2) ARTHROSCOPY PACK-LF, Pack Number DYNJ0390808F; 3) ARTHROSCOPY PACK-LF, Pack Number DYNJ23150J; 4) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 5) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 6) ARTHROSCOPY TRAY, Pack Number DYNJ46435A; 7) KNEE ARTHROSCOPY PACK, Pack Number DYNJ50120D
- Recall number
- Z-1452-2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2) ARTHROSCOPY PACK-LF, Pack Number DYNJ0390808F; 3) ARTHROSCOPY PACK-LF, Pack Number DYNJ23150J; 4) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 5) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 6) ARTHROSCOPY TRAY, Pack Number DYNJ46435A; 7) KNEE ARTHROSCOPY PACK, Pack Number DYNJ50120D
- Status
- Open, Classified
- Initiated
- 2025-02-11
- Posted
- 2025-03-25
- Terminated
- not on file
- Reason
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
- Root cause
- Under Investigation by firm
- Action
- Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
- Quantity
- 186 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Lot, serial or model codes
- 1) Pack Number DYNJHS0101I: UDI/DI 10888277046412 (each) 40888277046413 (case), Lot Number 21LBA780A; 2) Pack Number DYNJ0390808F: UDI/DI 10889942244591 (each) 40889942244592 (case), Lot Number 21JMG164A; 3) Pack Number DYNJ23150J: UDI/DI 10888277742239 (each) 40888277742230 (case), Lot Number 21LMF991A; 4) Pack Number DYNJ21233O: UDI/DI 10889942543878 (each) 40889942543879 (case), Lot Number 22GBQ106M; 5) Pack Number DYNJ21233O: UDI/DI 10889942543878 (each) 40889942543879 (case), Lot Number 22JBA123Z; 6) Pack Number DYNJ46435A: UDI/DI 10889942670901 (each) 40889942670902 (case), Lot Number 23LMG014A; 7) Pack Number DYNJ50120D: UDI/DI 10195327195656 (each) 40195327195657 (case), Lot Number 22JBG571Z
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OII
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28