ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031

Recall number
Z-1452-2020
Recalling firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Product
Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,
Status
Terminated
Initiated
2019-08-22
Posted
not on file
Terminated
2022-02-15
Reason
The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.
Root cause
Nonconforming Material/Component
Action
On 08/22/2019 a "Urgent - Field Safety Notice Medical Device correction" letter was sent to all affected consignees for the Concerto & Basic Shower Trolley. The notification informed consignees about the recall and asked they take the following action: 1. Verify the condition of all 4 Safety Catches attached to your device. If you notice the break withdraw the trolley from use. 2. IMPORTANT: The device may stay in use until the upgrade is performed, providing that the following instructions are carefully followed: - It is recommended that the Safety Catches on the shower trolley Side Supports be inspected daily to ensure any of the catches are not broken as show on the Image - The patient is not left unattended at any time. - The patient is fully assisted so that he/she does not grab, push or lean against the Side Supports during showering activities. - Caregivers should limit pulling and pushing of the Side Supports with excessive force. - Note: always adhere to the Safe Working Load as outlined in the IFU. 3. Ensure that all caregivers and users of Arjo Concerto & Basic shower trolleys are made aware of this Field Safety Notice (FSN) and all listed devices at your facility are available to be retrofitted free of charge during the service technician visit. 4. If you have purchased spare parts listed, any of the following will be done free of charge: a. If the spare parts were sold to a 3rd party, please provide the sold to information on the Customer Response Form. Arjo will contact the 3rd party to make arrangements for repair or return of the affected parts. b. If the purchased parts were installed on a Concerto/Basic unit in your inventory, please provide the serial number on the Customer Response Form so an Arjo technician can inspect and retrofit the unit. c. If the purchased parts are still in your inventory, please complete and return the Customer Response Form and Arjo will contact you and make arrangements to replace the spare
Quantity
669 trollys
Distribution
US: CA, CO, CT, GA, MA, MD, MI, NC, NJ, NY, OH, OK, TX, VA, WI OUS: Australia Austria, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iceland, India, Ireland, Italy, Japan, Korea, Luxembourg, Netherlands, Netherlands/CW, New Zealand, Norway, Philippines, Poland, Romaina, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, United Arab Emirates, United Kingdom
Lot, serial or model codes
Concert & Basic Shower trolleys Serial Numbers: P0512263 to P0551621, P0279164, SEE0643339, SEE0605124,P0249502, SEE045073, SEE0520032, 1106000338, SEE0348065, P0422951, P0232526
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
3007420694
Firm city
Komorniki
Firm state
not on file
Firm country
Poland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register