FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
- Recall number
- Z-1452-2015
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra Flowable Wound Matrix Size 3cc 1 unit/box single use, sterile device Rx Only Model Number FWD301 (US) & FDR301 (Non-US) Flowable Wound Matrix is comprised of the following components which are provided in a sealed and sterilized Tyvek / plastic tray: - An empty plastic syringe with Luer-Lok" tip - A plastic syringe with Luer-Lok" tip containing the Flowable Wound Matrix material - A syringe-to-syringe Luer-Lok" adapter / connector - A flexible plastic tube (injector) with Luer-Lok" connector Instructions for Use (IFU) and the sterilized package are subsequently packaged in a labeled and sealed box that is distributed to customers.
- Status
- Terminated
- Initiated
- 2015-03-13
- Posted
- 2015-04-16
- Terminated
- 2016-04-13
- Reason
- A lot of Flowable Wound Matrix that has been distributed to customers in late 2013 / early 2014 was verified as not meeting stability acceptance criteria for the attributes of visual appearance during dispensing and force vs. displacement (force needed for product to be extruded from the syringe).
- Root cause
- Unknown/Undetermined by firm
- Action
- Integra Life Sciences Inc. sent a Recall letter/Recall acknowledgement and return form dated March 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Integra asks customers to review inventory and determine if any of the affected inventory is in possession. The attached Recall Acknowledgement and Return Form should be completed and returned. Customer service will contact the customer upon receipt of the form and provide an RMA number. Questions can be directed to Customer Service at 1-888-301-0203.
- Quantity
- 287 packages ( 179 US 108 OUS )
- Distribution
- Worldwide Distribution - US Distribution and to the countries of and to the countries of Greece, France, Italy, Spain, Switzerland, South America, Turkey and United Kingdom.
- Lot, serial or model codes
- Model Number FWD301 (US) & FDR301 (Non-US) The following Lot Numbers on the packaged kits are affected: - 305000293250, 305000293251, 305000298616 (foreign) The syringes within the packaged kits are also Lot numbered. For the packaged kit Lot Numbers above, the corresponding respective syringe Lot Numbers are: - 10500028786, 105000288118, 105N00288965 (foreign)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KGN
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28