Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline procedure kits labeled as: 1) CSMC/ABDOMINAL HYST PACK-LF, Pack Number DYNJ0665086F; 2) MAJOR VAGINAL HARPER PACK-LF, Pack Number DYNJ0161768D; 3) ROBOT LAP TOTAL HYST, Pack Number DYNJ907811A; 4) VAG HYSTE PACK-LF, Pack Number DYNJ0773613J; 5) VAG HYSTERECTOMY PACK-LF, Pack Number DYNJ0549338Q

Recall number
Z-1449-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline procedure kits labeled as: 1) CSMC/ABDOMINAL HYST PACK-LF, Pack Number DYNJ0665086F; 2) MAJOR VAGINAL HARPER PACK-LF, Pack Number DYNJ0161768D; 3) ROBOT LAP TOTAL HYST, Pack Number DYNJ907811A; 4) VAG HYSTE PACK-LF, Pack Number DYNJ0773613J; 5) VAG HYSTERECTOMY PACK-LF, Pack Number DYNJ0549338Q
Status
Open, Classified
Initiated
2025-02-11
Posted
2025-03-25
Terminated
not on file
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Root cause
Under Investigation by firm
Action
Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
Quantity
268 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
1) Pack Number DYNJ0665086F: UDI/DI 10193489395204 (each) 40193489395205 (case), Lot Number 23JBW085A; 2) Pack Number DYNJ0161768D: UDI/DI 10195327107499 (each) 40195327107490 (case), Lot Number 22CDB491A; 3) Pack Number DYNJ907811A: UDI/DI 10193489984170 (each) 40193489984171 (case), Lot Number 21IBP928Z; 4) Pack Number DYNJ0773613J: UDI/DI 10193489863918 (each) 40193489863919 (case), Lot Number 21GMA889A; 5) Pack Number DYNJ0549338Q: UDI/DI 10195327092139 (each) 40195327092130 (case), Lot Number 23EMD924A
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJF
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register