FDA Medical Device Recalls (openFDA)
Roche Diagnostics Operations, Inc.: Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.
- Recall number
- Z-1449-2014
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.
- Status
- Terminated
- Initiated
- 2014-03-13
- Posted
- 2014-04-10
- Terminated
- 2018-01-10
- Reason
- The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c pack MULTI or empty pre-labeled cobas c packs. The ICVC feature allows the reagent probe to dive into the reagent cassette and confirm the reagent level in the bottles. This ensures
- Root cause
- Software Manufacturing/Software Deployment
- Action
- March 17, 2014 all consignees will be notified by Urgent Medical Device Correction sent via UPS Ground (receipt required). The notice identified the issue, root cause, affected assays with clinical significance. Consignees were instructed to follow the steps listed in the Urgent Medical Device Correction to correctly activate the ICVC and verify the correction. If your system is not connected to the Internet through the cobas link, and you do not have the option to run the affected reagents on an alternate system (i.e., cobas c 311 or cobas c 501 analyzer), please contact the Roche Support Network Customer Support Center at 1-800-428-2336 to have an update scheduled via a Roche Field Representative. If your system is connected to the Internet through the cobas link, follow the steps below: " Remove all reagent cobas c packs on board your cobas c 502 module for the assays listed on pages 1-2. " Dispose of these reagent cobas c packs according to your local waste management ordinances. " Load new cobas c packs for the affected assays, and visually verify the ICVC is performed. " Complete the attached fax form (5326-00-0314) and fax it to 1-888-276-6166. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " File this Urgent Medical Device Correction (UMDC) for future reference.
- Quantity
- 76 units
- Distribution
- USA Nationwide Distribution including San Juan PR
- Lot, serial or model codes
- not applicable
- 510(k) numbers
- ["K100853"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28