Skip to the content

FDA Medical Device Recalls (openFDA)

Biofire Defense: FilmArray NGDS Warrior Panel, Ref: NGDS-ASY-0007, UDI: (01)00851458005136, IVD, Rx Only

Recall number
Z-1448-2020
Recalling firm
Biofire Defense
Product
FilmArray NGDS Warrior Panel, Ref: NGDS-ASY-0007, UDI: (01)00851458005136, IVD, Rx Only
Status
Terminated
Initiated
2019-08-20
Posted
2020-02-18
Terminated
2020-10-30
Reason
The firm has identified that an incorrect instruction for use (IFU) revision was included in the qualitative, multiplexed, nucleic acid-based in vitro diagnostic test kit.
Root cause
Labeling Change Control
Action
On 8/20/2019, a "Urgent: Customer Recall Notice" letter was sent to all affected consignees informing them on the recall. In addition to providing information on the recall, the firm asked consignees do the following: 1. BFDf has enclosed a replacement printed copy of the Revision 02 Warrior IFU with this notice. BFDf advises customers to remove and dispose of all incorrect IFUs included in non-expired product from the affected lots listed. BFDf will provide additional replacement copies of the IFU on request. 2. Please complete and return the accompanying Recall Acknowledgement of Receipt Form included with this Notice, using one of the listed methods to confirm your receipt of this notification. 3. If you have any additional questions or concerns, please do not hesitate to contact BFDf Customer Support via email at Support@biofiredefense.com or via telephone at 1-801-262-3592 for product technical support.
Quantity
92 kits
Distribution
US: UT, MD, GA, CO, OH, TX, OUS: None
Lot, serial or model codes
Lot No./Expiration Date 457117D/2018-10-17 416118D/2019-03-15 439218D/2019-06-12 439318D/2019-07-18 454218D/2019-08-21 454318D/2019-08-22 455518D/2019-08-23 458118D/2019-09-18 458218D/2019-09-19 458018D/2019-09-20 464018D/2019-10-02 425719D/2020-05-30 462819D/2020-06-10
510(k) numbers
["K170883"]
PMA numbers
not on file
Product code
PRD
Firm FEI
3010770794
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register