Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories Inc.: ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate aminotransferse in human serum or plasma

Recall number
Z-1447-2012
Recalling firm
Abbott Laboratories Inc.
Product
ARCHITECT/AEROSET Activated Aspartate Aminotransferase The Activated Asparate Aminotransferase (Activated AST) assay is used for the quantitation of asparate aminotransferse in human serum or plasma
Status
Terminated
Initiated
2008-06-30
Posted
2012-04-23
Terminated
2012-04-23
Reason
This letter is to inform you that some samples may generate A-AST results of less than (<) 5 U/L or results within the reference range when actual concentration exceeds the non-Flex linearity of 1985 U/L. In other cases the following errors may be generated. -ARCHITECT cSystems: Error Code 1053 (Unable to calculate result, rate reaction linearity failure) -AEROSET: RL% Error Code Revised P
Root cause
Nonconforming Material/Component
Action
Abbott Laboratories sent a Product Correction letter dated August 5, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to do the following: Update your assay parameters for A-AST 1. Follow the steps in the appropriate attachment: - Use Attachment A for ARCHITECT cSystems - Use Attachment B for AEROSET. 2. All A-AST results of less than (<) 5 U/L should be evaluated by comparing to other clinically relevant assays Retain a copy of this letter for your laboratory records. For questions regarding this recall call 626-440-0705.
Quantity
1,646
Distribution
Worldwide Distribution including Germany
Lot, serial or model codes
List Numbers 8L91-20: 57028HW00, 59035HW00, 61031HW00, 64052HW00; List Numbers 8L91-40: 57017HW00, 59031HW00, 61038HW00, 64050HW00
510(k) numbers
["K071525"]
PMA numbers
not on file
Product code
CIT
Firm FEI
2018433
Firm city
South Pasadena
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register