Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer

Recall number
Z-1444-2020
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer
Status
Terminated
Initiated
2020-02-07
Posted
2020-03-05
Terminated
2024-05-30
Reason
The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.
Root cause
Nonconforming Material/Component
Action
On February 7, 2020, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers were informed of the product issue and given two options for CA 19-9 testing until another reagent lot is available for the cobas e 801 module: 1) Use a different Roche platform for testing, or 2) Implement a workaround, described in detail in the customer letter. The letter also provided instructions for receiving credit depending on which option was elected. Roche will notify customers when a new lot of CA 19-9 reagent for the cobas e 801 is available for ordering. ***Updated 10/5/20*** On 10/2/20, the firm issued an updated letter to customers to communicate an expansion to the recall. An additional lot was added to the recall scope.
Quantity
684
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IL, IN, KS, KY, MA, ME, MI, NC, NH, NJ, NY, TN, TX, WA and WI.
Lot, serial or model codes
Lot 416245 ***Updated 10/5/20*** 483123 ***Updated 8/5/22*** Additional released lots are subject to the same workaround: 46449, 50474,525510, and 568976
510(k) numbers
["K050231"]
PMA numbers
not on file
Product code
NIG
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register