Linvatec Corp.: ConMed Linvatec MicroChoice Footswitch and ConMed linvatec 3 Pedal Footswitch. Used to operate various surgical handpieces.

Recall number
Z-1444-06
Recalling firm
Linvatec Corp.
Product
ConMed Linvatec MicroChoice Footswitch and ConMed linvatec 3 Pedal Footswitch. Used to operate various surgical handpieces.
Status
Terminated
Initiated
2005-08-30
Posted
2006-08-29
Terminated
2012-07-13
Reason
The footswitch may develop a system fault resulting in the self-activation of the handpiece it controls, or it may cause the handpiece to run intermittently or to revert to a default setting.
Root cause
Other
Action
Linvatec has issued recall notification to all it''s direct accounts. The mailings were sent by UPS 08/31/2005. Each consignee was requested to return the affected product to Linvatec for a free replacement footswitch.
Quantity
118
Distribution
Worldwide, including United States and Canada.
Lot, serial or model codes
Serial numbers BBB81339-BBB81407, BBB91192-BBB91194, BBBB91209-BBB91366 and BBB91664-BBB91672.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GEY
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register