Skip to the content

FDA Medical Device Recalls (openFDA)

RESPIRATORY THERAPEUTICS GROUP LLC: RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.

Recall number
Z-1443-2020
Recalling firm
RESPIRATORY THERAPEUTICS GROUP LLC
Product
RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.
Status
Open, Classified
Initiated
2019-08-19
Posted
not on file
Terminated
not on file
Reason
It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
Root cause
Component change control
Action
On 8/19/19 a recall notice was sent to all affected customers. The letter outlined the recall issue and requested that customers complete the enclosed form for the return of the products (new and used) and the issuance of replacement product or credit.
Quantity
723 cases (7230 units)
Distribution
US Nationwide distribution in the states of CA, NJ, NY. OUS: None
Lot, serial or model codes
Lot KS1901013: 20019/03/31 - 06/192019 Lot KS1902002: 2019/06/21 - 07/31/2019
510(k) numbers
not on file
PMA numbers
not on file
Product code
BSY
Firm FEI
3011285025
Firm city
Los Angeles
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register