MEDLINE INDUSTRIES, LP - Northfield: Centurion manual surgical kits labeled as: a) ST. VIRTUS SPLINTER FORCEP 6" (SF8246), Product Code 66580; b) DEBRIDEMENT TRAY, Product Code I86630; c) BIOPSY INSTRUMENT PACK, Product Code I86690; d) DEBRIDEMENT TRAY, Product Code I86815; e) IR ADSON FORCEP & NEEDLE HOLDER, Product Code MNS12675; f) TOENAIL REMOVAL KIT, Product Code SUT20350 MEDLINE GENERAL PURPOSE INST. TRAY, Product Code DYNDA2346A
- Recall number
- Z-1442-2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Centurion manual surgical kits labeled as: a) ST. VIRTUS SPLINTER FORCEP 6" (SF8246), Product Code 66580; b) DEBRIDEMENT TRAY, Product Code I86630; c) BIOPSY INSTRUMENT PACK, Product Code I86690; d) DEBRIDEMENT TRAY, Product Code I86815; e) IR ADSON FORCEP & NEEDLE HOLDER, Product Code MNS12675; f) TOENAIL REMOVAL KIT, Product Code SUT20350 MEDLINE GENERAL PURPOSE INST. TRAY, Product Code DYNDA2346A
- Status
- Open, Classified
- Initiated
- 2024-02-23
- Posted
- 2024-03-29
- Terminated
- not on file
- Reason
- Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
- Root cause
- Process control
- Action
- Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.
- Quantity
- 7950 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Panama, Canada.
- Lot, serial or model codes
- a) 66580, UDI/DI 10653160028963 (case) 00653160028966 (each), Lot Numbers: 2023041090; b) I86630, UDI/DI 00653160286540 (case) 10653160286547 (each), Lot Numbers: 2023101190; c) I86690, UDI/DI 00653160290752 (case) 10653160290759 (each), Lot Numbers: 2023103190; d) I86815, UDI/DI 00653160342291 (case) 10653160342298 (each), Lot Numbers: 2023112190; e) MNS12675, UDI/DI 00653160337341 (case) 10653160337348 (each), Lot Numbers: 2023070390; f) SUT20350, UDI/DI 00653160306552 (case) 10653160306559 (each), Lot Numbers: 2023022290, 2023041190, 2023042490, 2023052390, 2023101890 MEDLINE: UDI/DI 20193489191582 (case) 10193489191585 (each), Lot Numbers: 2023062290, 2023081490
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KDD
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28