Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems Oncology Systems: Varian Medical Systems On-Board Imager, Versions 1.0.15 and 1.2.05 Imaging Accessory to Medical Linear Accelerator. The On-Board Imager Device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.

Recall number
Z-1442-2008
Recalling firm
Varian Medical Systems Oncology Systems
Product
Varian Medical Systems On-Board Imager, Versions 1.0.15 and 1.2.05 Imaging Accessory to Medical Linear Accelerator. The On-Board Imager Device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.
Status
Terminated
Initiated
2008-01-03
Posted
2008-08-15
Terminated
2010-11-27
Reason
If used with a third party radiation therapy treatment planning software system, mistreatment may occur because of a misalignment.
Root cause
Software design
Action
The firm has upgraded all user sites through normal course of service and installation activities. Notification letters were distributed starting January 03, 2008.
Quantity
291 units
Distribution
Worldwide Distribution -- USA including states of GA, MI, NY, MN, TX, TN, CA, MD, KS, AL, KY, CT, WI, LA, NC, NJ, PA, IL, RI, FL, NE, AK, VA, WV, NV, MO, TN, OR, IN, MA, CT, AL, AZ, NH, ID, OH, SC, CO, CT, ND, WA, and UT, and countries of Sweden, Switzerland, Germany, Denmark, Japan, UK, France, Italy, Norway, Taiwan, New Zealand, Canada, Poland, Australia, Colombia, China, Spain, Belgium, Russian Federation, India, Iceland.
Lot, serial or model codes
All serial numbers between H080002 to H080006, H080008 to H080014, H080016 to H080022, H080024 to H080050, H080052 to H080057, H080059 to H080066, H080068 to H080070, H080072 to H080075, H080077, H080078, H080080 to H080089, H080091, H080092, H080094 to H080105, H080107 to H080117, H080119 to H080134, H080136, H080137, H080139 to H080145, H080147 to H080158, H080160 to H080163, H080166, H080168 to H080200, H080202 to H080222, H080224 to H080228, H080230 to H080232, H080235, H080236, H080239, H080240, H080244, H080246 to H080248, H080251, H080254, H080256 to H080258, H080260 to H080263, H080265 to H080268, H080271 to H080282, H080286 to H080288, H080290 to H080292, H080294, H080296, H080301, H080303, H080304, H080306, H080307, H080309, H080311, H080312, H080314, H080318, H080320, H080321, H080324, H080325, H080328, H080329, H080335, H080348, H080351, H082000 to H082002, H082007, H082009 to H082012, H082020 to H082023, H082026, H082034, H082036, H082038, H082043, H082044, H082048, H082070, H082073, H082088, and H082204 .
510(k) numbers
["K040192","K041519","K042720"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register