Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Bionics, LLC: SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Recall number
Z-1441-2020
Recalling firm
Advanced Bionics, LLC
Product
SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.
Status
Terminated
Initiated
2019-04-04
Posted
not on file
Terminated
2022-06-27
Reason
The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.
Root cause
Software in the Use Environment
Action
On 4/4/2019, the recalling firm sent a letter to its customers via FedEx asking that the customer return the corrupted USB drive to the firm.
Quantity
364 versions of software
Distribution
US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI
Lot, serial or model codes
Lot #s:  663183 663188 663190 663858 663185 663184 663187 663186 664783 663860 664784 663859 669701 669100 669702
510(k) numbers
not on file
PMA numbers
["P960058"]
Product code
MCM
Firm FEI
1000220815
Firm city
Valencia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register