Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp: Anestar Anesthesia Delivery System

Recall number
Z-1441-04
Recalling firm
Datascope Corp
Product
Anestar Anesthesia Delivery System
Status
Terminated
Initiated
2003-06-24
Posted
2004-09-09
Terminated
2006-10-03
Reason
Failure of the Anestar AC Main Power Switch. Battery should last 30 minutes, then ventilation and monitoring will fail.
Root cause
Other
Action
A Field Correction action was initiated in June, 2003 and completed by the service reps in April, 2004.
Quantity
not on file
Distribution
The units were distributed to 29 hospitals nationwide. There are no US Government military or civilian customers.
Lot, serial or model codes
Serial#: 4600010001; 4600010003; 4600010004; 4600010007; 4600010008; 4600010009; 4600010010; 4600010012; 4600A20017; 4600A20018; 4600A20019; 4600A20020; 4600A20021; 4600A20023; 4600A20024; 4600A30066; 4600A30067; 4600A30068; 4600A30069; 4600A30070; 4600A30071; 4600A30072; 4600A30073; 4600A30074; 4600A30075; 4600B10002; 4600B20025; 4600B20026; 4600B20027; 4600B20028; 4600B20029; 4600B20030; 4600B20031; 4600B20032; 4600B20033; 4600B20034; 4600B30076; 4600B30077; 4600B30078; 4600B30079; 4600B30080; 4600B30081; 4600B30082; 4600B30083; 4600B30084; 4600B30086; 4600C30087; 4600C30088; 4600C30089; 4600C30091; 4600C30092; 4600C30093; 4600C30094; 4600C30095; 4600C30096; 4600C30097; 4600C30098; 4600C30099; 4600D20035; 4600D20036; 4600D20037; 4600D20038; 4600D20039; 4600D20040; 4600D20041; 4600D20042; 4600D20043; 4600D20044; 4600H20045; 4600H20046; 4600H20047; 4600H20048; 4600H20049; 4600H20050; 4600H20051; 4600H20052; 4600H20053; 4600H20054; 4600I10013; 4600J20055; 4600J20056; 4600J20057; 4600J20058; 4600J20059; 4600J20060; 4600J20061; 4600J20062; 4600J20063; 4600J20064.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BSZ
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register