Skip to the content

FDA Medical Device Recalls (openFDA)

Vyaire Medical: bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Recall number
Z-1439-2020
Recalling firm
Vyaire Medical
Product
bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.
Status
Terminated
Initiated
2019-11-25
Posted
not on file
Terminated
2023-08-08
Reason
The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.
Root cause
Software design
Action
Consignees were sent an Urgent Medical Device Recall letter beginning 11/25/2019 via overnight delivery. The letter identified affected devices, provided a description of the issue, and provided actions to be taken by distributors and end-users. A response form was asked to be returned. Questions can be directed to Vyaire's Technical Support Department at support.vent.us@vyaire.com.
Quantity
116
Distribution
US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.
Lot, serial or model codes
Serial numbers: MB170104, MB170118, MB170151, MB170202, MB230000, MB230020, MB230024, MB230027, MB230031, MB230033, MB230035, MB230036, MB230037, MB230038, MB230039, MB230040, MB230041, MB230046, MB230048, MB230049, MB230052, MB230053, MB230054, MB230055, MB230056, MB230057, MB230058, MB230059, MB230060, MB230061, MB230062, MB230063, MB230064, MB230065, MB230066, MB230067, MB230068, MB230069, MB230070, MB230071, MB230072, MB230073, MB230074, MB230075, MB230076, MB230077, MB230078, MB230079, MB230080, MB230081, MB230082, MB230083, MB230084, MB230085, MB230086, MB230087, MB230088, MB230089, MB230090, MB230091, MB230092, MB230093, MB230094, MB230095, MB230097, MB230098, MB230099, MB230100, MB230101, MB230102, MB230104, MB230190, MB230191, MB230194, MB230195, MB230196, MB230197, MB230201, MB230204, MB230205, MB230207, MB230209, MB230210, MB230217, MB230219, MB230220, MB230221, MB230222, MB230227, MB230228, MB230229, MB230230, MB230231, MB230232, MB230233, MB230243, MB230244, MB230246, MB230247, MB230249, MB230251, MB230253, MB230256, MB230258, MB230260, MB230261, MB230262, MB230263, MB230266, MB230270, MB230274, MB230275, MB230277, MB230294, MB230297, MB230298.
510(k) numbers
["K163127"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2050001
Firm city
Mettawa
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register