FDA Medical Device Recalls (openFDA)
Stryker Spine: Tritanium X TL Instruments - intervertebral fusion device with bone graft, lumbar. Catalogs:TL Inserter (580220100), TL Draw Rod (580220101), Tube Set/Syringe Adapter Assembly (680220100), Syringe Plunger (680220102).
- Recall number
- Z-1437-2021
- Recalling firm
- Stryker Spine
- Product
- Tritanium X TL Instruments - intervertebral fusion device with bone graft, lumbar. Catalogs:TL Inserter (580220100), TL Draw Rod (580220101), Tube Set/Syringe Adapter Assembly (680220100), Syringe Plunger (680220102).
- Status
- Terminated
- Initiated
- 2021-03-11
- Posted
- not on file
- Terminated
- 2023-01-05
- Reason
- Stryker identified non-conforming instruments that are components of the Tritanium X TL sets. The firm found non-conformances with the TL Inserter, TL Draw Rod, Tube Set/Syringe Adapter Assembly, and Syringe Plunger that could lead to device failure during a procedure.
- Root cause
- Device Design
- Action
- On 03/17/2021, Stryker issued an Urgent Medical Device Recall notice to customers and distributers via letter because the firm identified non-conforming components of the Titanium X TL sets which could lead to device failure in the field.
- Quantity
- 169 sets
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, GA, KY, MA, MT, NC, NE, NJ, NY, OH, PA, WY.
- Lot, serial or model codes
- (UDI): 07613327366310, 07613327366839680220102, 07613327366884, 07613327366457
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MAX
- Firm FEI
- 3004024955
- Firm city
- Allendale
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28