Fisher & Paykel Healthcare, Ltd.: Fisher & Paykel Healthcare ICON CPAP (continuous positive airway pressure). Models: ICON Auto: ICONAAN, ICONAAN-HT, ICONAAJ. ICON Novo: ICONNAN, ICONNAN-HT, ICONNAJ. ICON Premo: ICONPBN, ICONPBN-HT, ICONPBN-HTC. The ICON CPAP unit is for use on adult patients for the treatment of Obstructive Sleep Apnea (OSA). The device is for use in the home or sleep laboratory. The device is not intended to be used as a life supporting device.
- Recall number
- Z-1437-2013
- Recalling firm
- Fisher & Paykel Healthcare, Ltd.
- Product
- Fisher & Paykel Healthcare ICON CPAP (continuous positive airway pressure). Models: ICON Auto: ICONAAN, ICONAAN-HT, ICONAAJ. ICON Novo: ICONNAN, ICONNAN-HT, ICONNAJ. ICON Premo: ICONPBN, ICONPBN-HT, ICONPBN-HTC. The ICON CPAP unit is for use on adult patients for the treatment of Obstructive Sleep Apnea (OSA). The device is for use in the home or sleep laboratory. The device is not intended to be used as a life supporting device.
- Status
- Terminated
- Initiated
- 2013-04-22
- Posted
- 2013-05-30
- Terminated
- 2013-10-18
- Reason
- Fisher and Paykel Healthcare is recalling the ICON CPAP units because there is a potential power issue on certain ICON CPAP. These ICON CPAP units could potentially become inoperable, which would result in a lack of treatment to the patient's obstructive sleep apnea symptoms until a replacement unit is obtained.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Fisher & Paykel Healthcare Limited, sent a "Medical Device Recall" letter dated April 22, 2013 to all their customers who purchased the Fisher & Paykel Healthcare ICON CPAP Replacement. The recall letter described the product, notified the customers of the problem identified and the actions to be taken. Customers are instructed to locate and verify if any of the affected products are in stock; check the part and lot numbers of your device; complete and return the enclosed Product Recall Response form, even if you have not received or distributed any affected products via Email: FA-2013-001@fphcare.com; Fax: (949) 453-4001; or Post: Fisher & Paykel Healthcare,15365 Barranca Parkway, Irvine, CA 92618; transfer this notice immediately to all those who need to be aware and if affected product has been distributed to any other organizations or distributors, notify them regarding the recall within 10 business days upon receipt of this letter; prepare any products to be returned to Fisher & Paykel Healthcare and contact the field action coordinator via email at FA-2013-001@fphcare.com or directly at 1-800-792-3912 ext. 1426 to arrange for a return shipping label to be sent to you. If you have any questions related to this recall, contact the field action coordinator via email at FA-2013-001@fphcare.com or directly at (800) 792-3912 ext. 1426.
- Quantity
- 1,407 units
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Canada, Japan, and Taiwan.
- Lot, serial or model codes
- Lot 121210: December 10th, 2012. Lot 121221: December 21st, 2012.
- 510(k) numbers
- ["K094040"]
- PMA numbers
- not on file
- Product code
- BZD
- Firm FEI
- 3002806988
- Firm city
- North Shore City
- Firm state
- East Tamaki
- Firm country
- New Zealand
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28