Skip to the content

FDA Medical Device Recalls (openFDA)

Radiometer America Inc: TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)

Recall number
Z-1437-06
Recalling firm
Radiometer America Inc
Product
TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)
Status
Terminated
Initiated
2005-02-02
Posted
2006-08-30
Terminated
2012-05-09
Reason
Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.
Root cause
Other
Action
The firm initiated a recall action for the suspect device with a letter dated 2/2/2005. During a site visit to the U.S. customers conducted by the firm''s Field Service reps, the reps informed the user/customers of this problem and requested that all use of the device be discontinued, pending the formulation of a more permanent solution. Beginning on 2/8/2005, a second site visit was made to the same user/customers, at which time, the affected units were upgraded with operating software version 2.12. The firm''s recall/corrective action was scheduled for completion by 3/1/2005.
Quantity
9 monitors were distributed.
Distribution
Nationwide
Lot, serial or model codes
All Model #TCM4
510(k) numbers
not on file
PMA numbers
not on file
Product code
KLK
Firm FEI
1523456
Firm city
Westlake
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register