Heart Sync, Inc: Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Recall number
Z-1435-2013
Recalling firm
Heart Sync, Inc
Product
Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.
Status
Terminated
Initiated
2013-04-26
Posted
2013-05-30
Terminated
2014-04-16
Reason
On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.
Root cause
Process design
Action
HeartSync sent an URGENT MEDICAL DEVICE RECALL notification letter dated April 26, 2013, to all consignees of Heart Sync for C100-Physio Radiotransparent Defrillation Electrodes. On April 29, 2013, an URGENT MEDICAL DEVICE RECALL notification was sent to Vermed Inc for the D304 Defrillation Electrodes . The letters identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the affected product to the firm for full credit. Customers were asked to complete the form and send back via fax to 734-213-5640 or call 734-213-5530. For questions regarding this recall call 734-213-5530. .
Quantity
17180 units for all part numbers
Distribution
Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.
Lot, serial or model codes
Part Number: D304 and lot numbers:Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01
510(k) numbers
["K080421"]
PMA numbers
not on file
Product code
LDF
Firm FEI
1317188
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register