FDA Medical Device Recalls (openFDA)
Aspen Surgical Products, Inc.: 19FR Drain Kit Flat Full Perforated with Trocar, Saber 100ml Bulb and 3CSystem, Sterile (10/case) Intended for closed wound suction drainage
- Recall number
- Z-1433-2014
- Recalling firm
- Aspen Surgical Products, Inc.
- Product
- 19FR Drain Kit Flat Full Perforated with Trocar, Saber 100ml Bulb and 3CSystem, Sterile (10/case) Intended for closed wound suction drainage
- Status
- Terminated
- Initiated
- 2014-03-20
- Posted
- 2014-04-07
- Terminated
- 2014-11-25
- Reason
- The recall has been initiated due to a potential compromise of the sterile barrier of the Surgidyne Wound Drainage and Surgidyne Bulb Evacuator & Collection System medical devices.
- Root cause
- Packaging process control
- Action
- Aspen Surgical sent an Urgent Medical Device Recall letter dated March 20, 2014, to all affected customers and will be carried out at the distributor and retailer level, which can include doctor's offices, hospitals, and clinics. The distributor was instructed to contact their local FDA District office, conduct a sub-recall and to forward this Recall Notification and Response form to their retailers. A written notification to the consignees was sent by certified mail. Aspen has given the customers the option to either 1.) Return the product 2.) Destroy the product with a Certificate of Destruction to be provided to Aspen. Any returned product to Aspen will be put on MRB and dispositioned by Quality. The returned product will not be reprocessed. In order to determine effectiveness of the recall, any consignee that does not return the response form and/or product, will be contacted within three weeks of receipt of the Recall letter. Non-responders will be contacted a total of three (3) times in order to verify the effectiveness of the recall activity. Customers with questions were instructed to call 888-364-7004, ext 177 or 1-888-364-7004, ext. 156. For questions regarding this recall call 616-698-7100.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - USA including LA, MA, NH, TX, FL, OH, NC, PA, CA, UT, TN, MI, CO, WY, OK, WA, MN, PA, IN, VA, IA, IL, MD, CT, DE, SC, NY, AR, MA, AZ, NJ, MO, GA, KS, WV, NE, NC, Puerto Rico, and Internationally to Canada, UAE, England, Italy, Belgium, Costa Rica, and Ecuador.
- Lot, serial or model codes
- Product Number 340019
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GCY
- Firm FEI
- 3002750084
- Firm city
- Caledonia
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28