Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.

Recall number
Z-1433-04
Recalling firm
Beckman Coulter Inc
Product
COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.
Status
Terminated
Initiated
2004-08-12
Posted
2004-09-02
Terminated
2012-04-26
Reason
Beckman Coulter has confirmed an issue associated with the predilute mode of the COULTER LH 500 Hematology Analyzer. The workstation may display, transmit, and / or print an erroneous result or an incorrect dilution factor with a predilute sample result.
Root cause
Other
Action
Customers were notified via U.S. mail by a Product Corrective Action Letter (PCA) dated August 12, 2004. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occurring.
Quantity
222
Distribution
Nationwide & Canada
Lot, serial or model codes
Part Numbers: 178832, 178833, and 178834 software versions 1A and 1A2.
510(k) numbers
["K032000"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register