FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma.
- Recall number
- Z-1432-2012
- Recalling firm
- Beckman Coulter Inc.
- Product
- SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma.
- Status
- Terminated
- Initiated
- 2009-03-26
- Posted
- 2012-04-16
- Terminated
- 2012-04-18
- Reason
- The Ammonia Reagent lot M808278 may fail calibration or cause a low shift in Quality Control recovery on some SYNCHRON Systems.
- Root cause
- Unknown/Undetermined by firm
- Action
- Beckman Coulter mailed a Product Corrective Action (PCA) Letter on April 13, 2009 dated April 10, 2009 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue use and discard appropriately the lot number(s) of the affected product and contact their Customer Service Representative at (800) 526-3821 (US), or (800)463-7828 for reagent replacement. Customers were instructed to complete and return the enclosed response form within 10 days so that Beckman Coulter may maintain records. Customers were told to call the Customer Support Center (Hotline) at (800) 854-3633 with any questions regarding the PCA letter.
- Quantity
- 3,388 kits
- Distribution
- Worldwide Distribution - USA (nationwide) and the country of Canada
- Lot, serial or model codes
- Lot Number M808278
- 510(k) numbers
- ["K984402"]
- PMA numbers
- not on file
- Product code
- JIF
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28