Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma.

Recall number
Z-1432-2012
Recalling firm
Beckman Coulter Inc.
Product
SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma.
Status
Terminated
Initiated
2009-03-26
Posted
2012-04-16
Terminated
2012-04-18
Reason
The Ammonia Reagent lot M808278 may fail calibration or cause a low shift in Quality Control recovery on some SYNCHRON Systems.
Root cause
Unknown/Undetermined by firm
Action
Beckman Coulter mailed a Product Corrective Action (PCA) Letter on April 13, 2009 dated April 10, 2009 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue use and discard appropriately the lot number(s) of the affected product and contact their Customer Service Representative at (800) 526-3821 (US), or (800)463-7828 for reagent replacement. Customers were instructed to complete and return the enclosed response form within 10 days so that Beckman Coulter may maintain records. Customers were told to call the Customer Support Center (Hotline) at (800) 854-3633 with any questions regarding the PCA letter.
Quantity
3,388 kits
Distribution
Worldwide Distribution - USA (nationwide) and the country of Canada
Lot, serial or model codes
Lot Number M808278
510(k) numbers
["K984402"]
PMA numbers
not on file
Product code
JIF
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register