Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: namic convenience kits labeled as: KIT,ARTERIAL,60,(152CM),PG; medical convenience kit, REF 70036160

Recall number
Z-1429-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
namic convenience kits labeled as: KIT,ARTERIAL,60,(152CM),PG; medical convenience kit, REF 70036160
Status
Open, Classified
Initiated
2025-02-12
Posted
2025-03-21
Terminated
not on file
Reason
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Root cause
Process change control
Action
Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.
Quantity
40 units
Distribution
Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
Lot, serial or model codes
UDI/DI 10193489064803 (each), 20193489064800 (case), Lot number 0000140054
510(k) numbers
["K910764"]
PMA numbers
not on file
Product code
DRS
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register