Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Leksell Gamma Knife C

Recall number
Z-1429-2010
Recalling firm
Elekta, Inc.
Product
Leksell Gamma Knife C
Status
Terminated
Initiated
2006-08-01
Posted
2010-04-22
Terminated
2010-12-14
Reason
Due to a material change in a critical component of the linear actuators the quality of said part was found to be inadequate.
Root cause
Other
Action
Phase 1 - Technical Notes Leksell Gamma Knife C version 1.2 (dated August 2006) "Potential Problem with Error Handling" was sent to all users. Phase 2 - Mandatory Field Action 0000748, "Helmet Changer Actuator, MTB and MCU SW Upgrade Kit was released December 3, 2007. This Field Change Order includes two mod kits and a software upgrade to provide a permanent solution to the issues. Phase 3 - Technical Note 200 068, "Safety issue with helmet host actuator", released on May 15, 2008. This Technical Note includes a warning to customers of a potential quality problem with the helmet hoist actuator that is installed in the Leksell Gamma Knife. Phase 4 - Mandatory Field Action 200 067, "Investigation and Correction of Helmet Changer Actuators", released on May 15, 2008. This FCO includes a list of user machines which must be monitored for quality issues and provides a solution for replacement of parts should a deficiency be detected. Phase 5 - Mandatory Field Action 200 075, "Helmet Changer Actuator, MTB and MCU SW upgrade, release two", released on September 16, 2008. This FCO applies to machines within specific serial number ranges. Due to a material change in FCO-0000748 and baselines (conducted without notice by the manufacturer) the quality of said part was found to be inadequate. After reversing the material change, FCO 200 075 is released to update the installed base with the new type of actuator and to finish the remaining sites never or partially updated by the now superseded FCO-0000748. For further information, contact Elekta R&D Jonas Hallstensson on phone no. +46 709 843 553 or via e-mail jonas.hallstensson AT elekta.com.
Quantity
51 units
Distribution
AL, CA, CO, CT, FL, GA, IL, MD, MA, MI, MN, MO, NE, NY, NC, OH, OK, OR, PA, TN, TX, VA, WA
Lot, serial or model codes
4148, 4207, 4214, 4300, 4302, 4303, 4306, 4307, 4308, 4309, 4313, 4314, 4315, 4318, 4319, 4321, 4322, 4326, 4327, 4328, 4329, 4332, 4333, 4334, 4337, 4338, 4339, 4340, 4341, 4342, 4343, 4344, 4346, 4351, 4352, 4353, 4354, 4357, 4363, 4365, 4366, 5000, 5003, 5006, 5009, 5010, 5018, 5019, 5022, 5023, 5026
510(k) numbers
["K984328"]
PMA numbers
not on file
Product code
IWB
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register