Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Panta Arthrodesis Nail Support device, 510 (k) #K050882. One (1) unit of product is kitted within an instrumentation tray. The product is available in six (6) sizes. All nails are color-coded for easy size identification. Bony fixation is achieved using two (2) tibial screws, two (2) calcaneal screws and one (optional) talar screw. The nail, cross-locking screws and end cap are manufactured from a titanium alloy: TI-6AI-4V ELI ISO 5832-3, ASTM F136. The PANTA Arthrodesis Nail is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. The Panta Nail is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include: " Post-traumatic and degenerative arthritis involving both ankle and subtalar joints. " Rheumatoid arthritis. " Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body. " Revision of failed total ankle arthroplasty with subtalar intrusion. " Talar deficiency conditions (requiring a tibiocalcaneal arthrodesis). " Avascular necrosis of the talus. " Neuroarthropathy or neuropathic ankle deformity. " Severe deformity as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease. " Severe pilon fractures with trauma to the subtalar joint.

Recall number
Z-1428-2012
Recalling firm
Integra LifeSciences Corp.
Product
Panta Arthrodesis Nail Support device, 510 (k) #K050882. One (1) unit of product is kitted within an instrumentation tray. The product is available in six (6) sizes. All nails are color-coded for easy size identification. Bony fixation is achieved using two (2) tibial screws, two (2) calcaneal screws and one (optional) talar screw. The nail, cross-locking screws and end cap are manufactured from a titanium alloy: TI-6AI-4V ELI ISO 5832-3, ASTM F136. The PANTA Arthrodesis Nail is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. The Panta Nail is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include: " Post-traumatic and degenerative arthritis involving both ankle and subtalar joints. " Rheumatoid arthritis. " Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body. " Revision of failed total ankle arthroplasty with subtalar intrusion. " Talar deficiency conditions (requiring a tibiocalcaneal arthrodesis). " Avascular necrosis of the talus. " Neuroarthropathy or neuropathic ankle deformity. " Severe deformity as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease. " Severe pilon fractures with trauma to the subtalar joint.
Status
Terminated
Initiated
2011-05-19
Posted
2012-04-16
Terminated
2013-02-19
Reason
The firm received multiple complaints from their user customers (surgeons) who reported encountering difficulty inserting a calcaneal or tibial screw through the holes of the PANTA nail.
Root cause
Use error
Action
Integra issued a Field Safety Notice with an attached Acknowledgement and return Form to all Integra Sales Representatives and Customers via Fed Ex and E-mail on May 19, 2011. The Field Safety Notice informs the customers of the firm's receipt of information concerning several reported incidents that lead to the misalignment of the calcaneal of tibial when used with the Panta Arthrodesis Nail Support Device. The notice goes stated that Integra determined that the misalignment of these screws may be attributed to improper surgical technique and to that end, the Field Safety Notice is being issued in order to reinforce the importance of following the recommended guidelines our lined in the product's Surgical Technique Manual (a copy of the referenced guidelines are provided with the Field Safety Notice (FSN). The FSN goes on to direct the reader/customer to applicable section sections of the manual to which particular attention should be directed in order to minimize the potential for misalignment of the screws. The customers are asked to distribute the information provided to the surgical teams at their various institutions. Moreover, they are also asked to sign and return the enclosed Field Safety Notice Acknowledgement and Return Form so as to document their receipt of the notification and the material provided. Should the customer(s) have questions, they are directed to contact Sean Luland, Director of Marketing-Lower Extremity Reconstruction at Integra LifeScience at: 513-533-7929, for additional information. For questions regarding this recall call 609-275-0500.
Quantity
64 units were distributed
Distribution
Nationwide Distribution including MD, IL, NU, CA, GA, CO, OH, SD, AR, FL, TX, MA, MO, NC, OK, KY, NJ, UT, VA, MN, DC, PA, WA, CT, AZ, IL, MI, AL, NE, TN, NM, NV, LA, OR, DE, WI, ND, NH, MS, SC, IN and Puerto Rico.
Lot, serial or model codes
Model #519110, Batch Numbers: E5PS, E7D8, E6QU/1, E1LV, E43W, E1LV/1, E698, E1RR, E43W/1, EJZR, E1LQ, E698/1, and E7D8/1
510(k) numbers
["K050882"]
PMA numbers
not on file
Product code
HSB
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register