FDA Medical Device Recalls (openFDA)
Cook Incorporated: Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8.5 (G57692)
- Recall number
- Z-1427-2023
- Recalling firm
- Cook Incorporated
- Product
- Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8.5 (G57692)
- Status
- Open, Classified
- Initiated
- 2023-04-07
- Posted
- 2023-05-10
- Terminated
- not on file
- Reason
- There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
- Root cause
- Nonconforming Material/Component
- Action
- An URGENT: MEDICAL DEVICE RECALL dated 4/7/23 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets and Trays from the Cook Medical lots in Attachment A. If so, quarantine any affected sets/trays that remain unused. Immediately cease all further distribution and use of the affected sets/trays. 2. Return the complete sets/trays to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected sets/trays on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected sets/trays have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market and replacing potentially defective tracheostomy tubes with new tracheostomy tubes as they become available from Medtronic. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your return of the Acknowledgement and Receipt Form.
- Quantity
- 3.053 units
- Distribution
- Worldwide distribution - US Nationwide.
- Lot, serial or model codes
- C-PTIS-100-HC-G-EU-FLEX7.5 (G57695): UDI-DI 00827002576957, Lot Number 14842940; C-PTIS-100-HC-G-EU-FLEX8.5 (G57696): UDI-DI 00827002576964, Lot Numbers 14727192, 14332964, 14335896, 14380758, 14380755, 14332978, 14332973, 14335892, 14335893, 14464000, 14332985, 14464008, 14463988, 14503228, 14503224, 14503222, 14503229, 14503227, 14742869, 14742870, 14727207, 14742873; C-PTIS-100-HC-G-NA-FLEX7.5 (G57691): UDI-DI 00827002576919, Lot Numbers 15063922, 15063923, 15103482, 14227740, 14221180, 14204927, 14227743, 14221125, 14221198, 14221190, 14163328, 14250265, 14250269, 14317245, 14317213, 14345634, 14376893, 14633088, 14633089, 14794549, 14794548, 14794547, 14874486, 14874484, 14926500, 14967737, 14926505, 14967739, 14967741, 14221180; C-PTIS-100-HC-G-NA-FLEX8.5 (G57692): UDI-DI 00827002576926, Lot Numbers 14608362, 14750607, 14894816, 14264399, 14299799, 14299802, 14299803, 14345667, 14365693, 14365699, 14365696, 14365694, 14365701, 14345687, 14345587, 14456438, 14456443, 14456431, 14407846, 14750610, 14488394, 14488418, 14608361, 14608363, 14750618, 14608360, 14961742, 14961716, 15089037, 15089038, 15136113
- 510(k) numbers
- ["K193133"]
- PMA numbers
- not on file
- Product code
- JOH
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28