Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Hudson RCI, BiteGard Oral Bite Block, Single Patient Use, Rx Only, Teleflex Medical, Research Triangle Park, NC. The Teleflex Medical BiteGard Molar Bite Black is a disposable single patient use device that fits between the patients molars to prevent the patient from closing their mouth entirely and from obstructing the Laryngeal Mask Airway (LMA) tube, in the case where one is used.

Recall number
Z-1427-2012
Recalling firm
Teleflex Medical
Product
Hudson RCI, BiteGard Oral Bite Block, Single Patient Use, Rx Only, Teleflex Medical, Research Triangle Park, NC. The Teleflex Medical BiteGard Molar Bite Black is a disposable single patient use device that fits between the patients molars to prevent the patient from closing their mouth entirely and from obstructing the Laryngeal Mask Airway (LMA) tube, in the case where one is used.
Status
Terminated
Initiated
2012-04-03
Posted
2012-04-16
Terminated
2014-05-20
Reason
Teleflex Medical received complaints that the bite block can become separated from the handle. If separation occurs, medical intervention may be required to remove the bite block from the patient's trachea or esophagus/GI tract.
Root cause
Process control
Action
Consignees received "Urgent Medical Device Recall Notification" letters on or about 04/03/2011. The letters described the issue and provided recommended actions regarding remaining product and returns. Questions were directed to Customer Service at (866) 804-9881 (US).
Quantity
333,724 units
Distribution
Worldwide Distribution, including Nationwide (USA) and the countries of Canada, Singapore, China, Denmark, Netherlands, India, Australia, Japan and the United Kingdom.
Lot, serial or model codes
Catalog number: 1140; Lot numbers: 02A0800905, 02C1100483, 02F1000054, 02J0902663, 02A0801994, 02C1101954, 02F1000055, 02J1000939, 02A1000256, 02C1102754, 02F1100700, 02J1001824, 02A1000257, 02D0800059, 02F1101638, 02J1001825, 02A1001314, 02D0801329, 02G0801052, 02K0801935, 02A1001766, 02D0900557, 02G0802642, 02K0803847, 02A1100326, 02D0901732, 02G0900707, 02K0803848, 02A1101331, 02D1100098, 02G0901295, 02K0803849, 02A1102273, 02E0800110, 02G0902610, 02K0900029, 02B0900298, 02E0800342, 02G1000387, 02K0901174, 02B1000808, 02E0802363, 02G1000547, 02K1000019, 02B1000809, 02E0900148, 02G1100211, 02K1000520, 02B1001744, 02E0900638, 02H0800052, 02L1000392, 02B1100389, 02E1000111, 02H0800366, 02L1002223, 02B1101712, 02E1001758, 02H0801439, 02M0900100, 02C0800010, 02E1002588, 02H0801440, 02M0900855, 02C0802808, 02E1100223, 02H0900911, 02M0901463, 02C0901197, 02E1101181, 02J0800358, 02M0902072, 02C1001856, 02F0801932, 02J0801231, 02M1000288, 02C1001857, 02F0900762, 02J0900815 and 02M1000824 and Catalog number: 41140 (sold internationally) Lot numbers: 02A1102347, 02C1000226, 02C1003651, 02C1100970, 02D1001954, 02D1002705, 02D1101766, 02E1000107, 02E1100311, 02G0900700, 02J0802603, 02J1001026, 02K0901769, 02K1000616, 02K1001588, 02L0901972 and 02L1000486.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KOG
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register