Skip to the content

FDA Medical Device Recalls (openFDA)

SPINEART SA: JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S

Recall number
Z-1426-2023
Recalling firm
SPINEART SA
Product
JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S
Status
Open, Classified
Initiated
2023-03-16
Posted
2023-04-24
Terminated
not on file
Reason
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
Root cause
Process change control
Action
On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to orders@spineart.us Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at orders@spineart.us
Quantity
544
Distribution
US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
Lot, serial or model codes
REF/UDI-DI/Lot: JLT-PL 02 08-S/07640178987388/4-8826, 5-4142, 5-5377, 5-0643, 5-2431; JLT-PL 02 10-S/07640178987395/6-5488, 6-6556, 5-0644, 6-4734, 6-4735, 5-5378, 6-3211; JLT-PL 02 12-S/07640178987401/6-5489, 6-6558, 6-7645, 6-8734, 6-4732, 4-8828, 5-5379, 6-4733; JLT-PL 02 14-S/07640178987418/5-8439, 4-8829, 6-0714, 5-5380; JLT-PL 02 16-S/07640178987425/4-8830, 5-4145, 6-3217, 6-0726, 6-5490; JLT-PL 04 08-S/07640178987456/4-8845, 5-4146, 6-0725; JLT-PL 04 10-S/07640178987463/4-8831, 5-4147, 6-0722; JLT-PL 04 12-S/07640178987470/4-8832, 5-4148, 5-8438, 6-0721, 6-3209; JLT-PL 04 14-S/07640178987487/4-3861, 4-8833, 5-4149; JLT-PL 04 16-S/07640178987494/5-4150; JLT-PL 04 18-S/07640178987500/5-4151
510(k) numbers
["K190877"]
PMA numbers
not on file
Product code
OVD
Firm FEI
3007728266
Firm city
Plan-Les-Ouates
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06