Skip to the content

FDA Medical Device Recalls (openFDA)

Richard Wolf GmbH: FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685

Recall number
Z-1424-2023
Recalling firm
Richard Wolf GmbH
Product
FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685
Status
Open, Classified
Initiated
2023-03-24
Posted
2023-04-24
Terminated
not on file
Reason
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
Root cause
Under Investigation by firm
Action
The firm issued an Urgent Medical Device Correction notice on 03/24/2023 to its consignees by USPS certified mail. The notice explained the issue and explained the proper use of the device to avoid exposing patients to improper handling. Richard Wolf is already optimizing the forceps design. Further batches of these forceps can be used with the expected maximum flexibility. The procedure described in the notice does not expose patients to risks from jaw sections that do not open. The firm issued a follow up letter titled, Urgent Medical Device Removal beginning on 08/02/2023, via USPS certified mail. The notice advised customers that the device has been redesigned. Customers were instructed to review inventory, ensure all users are informed of the recall, notify customers if further distributed, return the customer response form, arrange for the shipment of revised forceps and return affected devices. The firm issued an updated letter to customers who have not yet responded to the original notification on about 09/12/2023. The updated letter included an image of where to locate the lot/batch number on the forceps, and the table of lot/batch numbers was also reorganized for better readability.
Quantity
319 units
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czech Republic (CZ), Germany (DE), Denmark (DK), Ecuador (EC), Spain (ES), Finland (FI), France (FR), Greece (GR), Hong Kong (HK), Croatia (HR), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), South Korea (KR), Kuwait (KW), Kazakhstan (KZ), LB, Lithuania (LT), Marzocco (MA), Mexico (MX), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Pakistan (PK), Poland (PL), Palestine (PS), Portugal (PT), Qatar (QA), Romania (RO), Serbia (RS), Russian Federation (RU), Saudi Arabia (SA), Slovenia (SI), Slovakia (SK), Thailand (TH), Turkey (TR).
Lot, serial or model codes
UDI/DI: 04055207020265; Lot Numbers: 4500377618, 4500377296, 4500374239, 4500371819, 4500367415, 4500363102, 4500367414, 4500363666, 4500363102, 4500359744, 4500357831, 4500354319, 4500352415, 4500349678, 4500344619, 4500347121, 4500349678, 4500348605, 4500344619, 4500347121, 4500344619, 4500341300, 4500338554, 4500337773, 4500335110
510(k) numbers
not on file
PMA numbers
not on file
Product code
FCL
Firm FEI
3002808001
Firm city
Knittlingen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.