Randox Laboratories Ltd.: Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD

Recall number
Z-1423-2022
Recalling firm
Randox Laboratories Ltd.
Product
Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD
Status
Terminated
Initiated
2022-06-09
Posted
not on file
Terminated
2024-02-08
Reason
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
Root cause
Under Investigation by firm
Action
On June 9, 2022, the firm notified customers of the recall via email. The issue only affects Siemens Atellica CH analyzer users. No action is required if the assay is being run on other platforms. The issue will be resolved in Atellica Solution Software (SW) version 1.25.3. Until all Atellica CH analyzers in the laboratory are updated to SW 1.25.3, the assay can be run by Open Channel. Please contact applications@randox.com for an Atellica CH Open Channel application sheet. If the assay is not run by Open Channel prior to installation of SW version 1.25.3, the assay must be re-calibrated once 1.25.3 is installed. Customers were instructed to review results generated with the affected batches in line with the clinical profile of the patient.
Quantity
12 kits (US)
Distribution
US Nationwide distribution in the state of IL.
Lot, serial or model codes
UDI-DI (GTIN): 05055273215557 All lots within expiry
510(k) numbers
["K002585"]
PMA numbers
not on file
Product code
DDG
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register