FDA Medical Device Recalls (openFDA)
St. Jude Medical: EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575 Route 73 North, Units D, West Berlin, NJ 08091. Indicated for use during clinical electrophysiology procedures.
- Recall number
- Z-1422-2009
- Recalling firm
- St. Jude Medical
- Product
- EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575 Route 73 North, Units D, West Berlin, NJ 08091. Indicated for use during clinical electrophysiology procedures.
- Status
- Terminated
- Initiated
- 2009-03-23
- Posted
- 2009-06-12
- Terminated
- 2010-01-20
- Reason
- Non-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 calls for both lines to be fused.
- Root cause
- Device Design
- Action
- All foreign customers were sent Urgent Product Safety Information letters on March 23, 2009. Letters described units affected, the reason for action, and the actions to take to eliminate risk. It is recommended that the information in the letter be circulated to all users, risk managers and the biomedical engineering department responsible for maintaining and inspecting electrical equipment located within their institution. The letter also stated that a SJM representative will be in touch to perform the EP-4 Stimulator fuse modification. Customers are to complete and return the attachment.
- Quantity
- 694 total (US, Canada, and OUS)
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- Systems Part Numbers: EP-4I-2-110; EP-4I-2-220; EP-4I-4-110; EP-4-220: WM-01M-241; WM-01M-560; WM-01M-1200; WM-02M-241; WM-02M-560; and WM-02M-1200.
- 510(k) numbers
- ["K040207","K994011"]
- PMA numbers
- not on file
- Product code
- JOQ
- Firm FEI
- 2126673
- Firm city
- West Berlin
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28